The
inspection carried out by the State consists of verifying that
the instruments meet the following regulatory requirements:
Verification of three points at the top, center and bottom
of the scale.
The maximum error tolerated must not be more than:
- 0.1 % vol. spread throughout the length of the scale for the
alcoholometers.
- 0.3 g/dm3 spread throughout the length of the scale for hydrometers.
The
products conformity is attested:
By the engraving of the inspection date, name of the
state laboratory that carried out the testing and a mark guaranteeing
the positioning of the scale.
By a verification report.
 |
|

Verification Report
|
The
instruments can be COFRAC certified. In this case, they are
calibrated by hydrostatic weighing with continuous variation
in the depth of the instruments immersion in the reference
liquid kept at a temperature of +20°C ± 0.08°C
(CUCKOW method). The inspection certificate that accompanies
the instrument is complete. It gives the real values, the instruments
indications and the right corrections to be applied.
|
|
|
|
 |
|
Inspection
Certificate
|
Alcoholometers
are graduated for a maximum range of 10% vol; accuracy 0.1 %
vol. They are available in a case of 5 or 10 accompanied by
a thermometer.
Standard range 0/10, 10/20
90/100.
Specials 6/16, 45/55 - 95/103 - 9/16, 11/18, 15/22.
Hydrometers
are graduated for a maximum range of 30g/dm3; accuracy 0.2g/dm3.
Standard range 970/1000, 1000/1030, 1030/1060
1120/1150 and 990/1020
Special (983/985)/1003 also called Extracto-oenometer.
Accessories:
the thermometers are also inspected (bearing a specific marking)
and the test tubes are calibrated.

The
thermometers are graduated in ° Celsius and placed in a
protective glass tube.
Range 0/35 in 1/2°, 0/50 in 1/2°, -20/+35 in 1/2°,
-1/+31 in 1/10°.
|